Leave Your Message
Hydrophobic Transducer Protector Filter With Micropore PTFE
Transducer Protector Filters

Hydrophobic Transducer Protector Filter With Micropore PTFE

Category: T
Filter Media: Hydrophobic PTFE 
Housing: ABS-based polymers
Removal of Particles options:  0.22/0.45/1.0μm
Sterilization:Ethlene oxide (Max 60℃) 
Standard: GB/T1962.2-2001,ISO 594-2 in compliance with ISO9001 

    Product Introduction

    Transducer Protector Filter with hydrophobic PTFE membrane(0.22/0.45/1.0μm) in an ABS-based polymers housing. Non-toxic, non-sterile. The product is provided for further manufacturing, processing and repackaging. PTFE filter membranes adopt a dual-layer lamination design, which offers more comprehensive filtration effects. Air entering from any direction is effectively treated with antibacterial filtration while ensuring venting.Hemodialysis machine's circuits are equipped with Transducer Protectors Filters (TP filters) to protect the equipment sensor and reduce the risk of infection control to some extent.


    Features

    • Equipped hydrophobic PTFE
    • Antibacterial and air venting
    • Allows only sterile air to pass through
    • Protect patients and equipment from both bacterial and viral cross-contamination
    • Eliminates contamination of the internal pressure monitoring lines
    • Prevents blood and dialyzing fluid from damaging sensitive transducers
    TP filter in hemodialysis - 2u7y

    Specifications

    Dimensions

    25.5*21.2mm (Luer connector included) 
    Connector Luer lock connector compatible with GB/T 1962.2-2001(ISO 594-2)
    Max burs pressuret  4.0bar (58.0 psi) 
    Venting area

    1.13cm² 

    Operating temp 10-30℃ 
    Pyrogenicity ≤0.5EU/PC
    Storage temp 5-40℃ 
    Sterilization Ethlene oxide(Max 60℃) 

    Perfomance

    tp-ventww8

    Applications

    ● Hemodialysis 

    Transducer protectors filters are used in hemodialysis blood lines, one each for arterial and venous pressures. TP filters with hydrophobic filtration membranes can separate the liquid (priming fluid, blood) in the circuit from the transducer, preventing the blood flowing through the circuit from contaminating the machine.

    ● In-line air venting filter for Intensive Care Unit
    Transducer protector filters guarantee a a sufficient air flow to provide immediate reaction time if there is a change in patient blood pressure. .

    package

    6000pcs/box, 3 inner PE bags with 2000 pcs inside each. each bag is sealed separately and cushion between each other.

    Each bag is labeled with information as follwing, differrent lots in one box are closed separately and labeled seperately to prevent mix up. 

    labeled information

    Name

    Model

    Specifications

    Quantity

    Batch no.

    Operator no.

    warnings

    ● Filter for medical use, assembled in clean room only. Remove the external bag before planting into a clean room. 
    ● Handle with care. 
    ● Alcohol is not allowed to contact with housing, or it will damage the ABS-based housing visible or invisible. 
    ● It's not recommended to use alcohol or any kind of disinfactant in direct contact with filter media, PTFE membrane will become hydrophilic until dried.

    Ordering Information

    Apperance Housing Filter media options Mold Connection  Vents
    GLXL
    012
    022 T 2Y N
    GLXY
    045
    GLXY
    100