Leave Your Message
China XINNA F33 IV Filter 1.2/2.0/3.0/5.0μm – Quality Infusion Set Filter from Leading Suppliers and Factory
Hot Products

China XINNA F33 IV Filter 1.2/2.0/3.0/5.0μm – Quality Infusion Set Filter from Leading Suppliers and Factory

Discover our premium-quality products sourced directly from trusted China suppliers. Our factory utilizes state-of-the-art manufacturing techniques to ensure superior craftsmanship and durability. We are dedicated to providing exceptional value and reliability, making us the go-to choice for businesses seeking high-quality goods. With our commitment to excellence, you can trust that you are partnering with one of the leading suppliers in the industry. Enhance your product lineup with our offerings today!
  • Category F
  • Filter Membrane Hydrophilic Polypropylene
  • Vent Hydrophobic PTFE membrane
  • Housing mAbs
  • Removal of Particles options 1.2/2.0/3.0/5.0μm
  • Standard YY0286.1 GB8368-2018 ISO 8536-4

Product Introduction

Potential risks caused by harmful particles, microbial contamination, and air micro embolism should be considered in infusion therapy, drug preparation, and mixing processes. F33 IV filter is designed for use in filtration of intravenous for removal of particles large than 1.2/2.0/3.0/5.0μm.

F33 IV Filter is an air eliminating filter for the removal of inadvertent particulate debris and other components like incomplete mixing of drugs and entrained air in catheter which may be found in solutions intended for intravenous use. The filtration efficiency of the membrane and the overall design of the filter can impact the safety of drug delivery for human beings.

It is inbuilt with low binding Hydrophilic Polypropylene filter membrane lower the loss of drug and active ingredients, and dual vents design allows air elimination in any position, ensuring the safety and effectiveness of intravenous.

Features

  • Range of pore size options
  • Low protein binding & Low dissolution
  • Dual vent design ensures quick Priming and air elimination
  • Compatibility with TPN & chemotherapeutic agents
  • Maximizes filtration area with compact construction
  • Lightproof-colored housings for light sensitive drugs is available
  • Printing available for ease of product identification
in-line infusion filter in PES with 2 window PTFE vents

Characteristics

Parameter Specification
Dimensions 57*29*0.94mm (inlet/outlet connector included)
Connector Φ 3.1mm/3.6mm
Max burst pressure 3.6bar (52.5psi)
Priming volume ≤2.0ml
Venting air flow rate 25ml/min@5kpa
Operating temp 10-30℃
Pyrogenicity ≤0.5EU/ml
Storage temp 5-40℃
Sterilization Ethlene oxide (Max 60℃)

Performance

F33 Performance Chart

Package

3000pcs/carton, 4 inner PE bags with 750 pcs inside each. Each bag is sealed separately and cushion between each other.

Each bag is labeled with information as following, different lots in one box are closed separately and labeled seperately to prevent mix up.

Label information

Name Specifications Batach No.
Model Quantity Operator No.

Warnings

● Filter for medical use, assembled in clean room only. Remove the external bag before planting into a clean room.
● Handle with care. Keep filter in vertical position with flow arrow on both sides and the printing mark before assembling. Be careful on glueing, if Hydrophilic PES membrane contact with glue, membrane will breaks down.
● Alcohol is not allowed to contact with housing, or it will damage the ABS-based housing visible or invisible.
● It's not recommended to use alcohol or any kind of disinfactant in direct contact with venting membrane, PTFE membrane will become hydrophilic until dried.
● Blood delivery is not allowed, the filter clogs with blood or red blood cells. Filters with 0.2μm filter media is not recommended for TPN and lipids administration.
● Rigorous testing is recommended before marketing if drugs is not common.
● Contact with Xinna Medical for more information if any doubt is exist advanced.

Ordering information

Appearance Housing Filter media options Mold Connector Vents
GLOL 033 500 F OY Y
GLBC 300
GLBZ 200
120

Frequently Asked Questions

What is the primary function of the F33 IV Filter?

The F33 IV Filter is designed to filter intravenous solutions, removing harmful particles larger than 1.2/2.0/3.0/5.0μm, incomplete drug mixtures, and entrained air to prevent potential risks like microbial contamination and air micro embolism.

How does the filter minimize drug loss during infusion?

It is built with a low-binding Hydrophilic Polypropylene filter membrane. This configuration reduces the absorption of drugs and active ingredients, ensuring the dosage delivery remains safe and effective.

Can the F33 IV Filter be used for blood transfusions?

No, blood delivery is not allowed because the filter will clog with blood or red blood cells. Additionally, filters with 0.2μm filter media are not recommended for TPN and lipids administration.

What chemical contacts should be avoided during assembly and use?

Avoid contacting the ABS-based housing with alcohol to prevent visible or invisible damage. Additionally, keep glue away from the Hydrophilic PES membrane to prevent structural breakdown, and avoid direct contact of alcohol/disinfectants with the PTFE venting membrane.

What are the storage and packaging specifications for the F33 IV Filter?

The filters are packaged 3000pcs per carton (4 inner PE bags with 750pcs each, individually sealed and cushioned). They should be stored at a temperature range of 5-40℃, and the operating temperature range is 10-30℃.