Leave Your Message
0.2μm Hemodialysis Blood Line Transducer Protectors - China Suppliers & Factory Solutions
Hot Products

0.2μm Hemodialysis Blood Line Transducer Protectors - China Suppliers & Factory Solutions

Discover high-quality products directly from China’s leading suppliers and factory. Our manufacturing facility combines advanced technology with skilled craftsmanship to ensure top-notch quality and competitive pricing. As one of the trusted China suppliers, we specialize in delivering reliable and customized solutions to meet diverse market needs. Partner with our factory to benefit from efficient production, strict quality control, and timely delivery. Whether you’re sourcing in bulk or require tailored products, our China factory and suppliers are committed to supporting your business growth with excellence and professionalism
  • Category TP
  • Filter Media PTFE
  • Pore Size 0.2 μm
  • Housing mAbs
  • Standard GB/T 1962.2-2001
  • ISO 594-2 ISO9001
  • Intended Use Prevent blood and dialyzing fluid from damaging pressure transducer in hemodialysis

Product Introduction

Transducer Protector is an antibacterial hydrophobic filter primarily designed to serve as an inline protective medical device for pressure monitors within the hemodialysis system. With a 0.22μm PTFE membrane filter built-in, transducer protector allows only sterile air to pass through and meanwhile works as a good protective barrier to keep the blood and dialyzing fluid of the circuit separated from the dialysis machine, preventing contamination and damage to the pressure monitoring system, and protecting patients and equipment from cross contamination. It ensures the accuracy and reliability of the sensitive transducers and the entire monitoring system, thus improving the treatment outcomes and safety of patients.

Features

  • 0.2 μm hydrophobic filter
  • Female luer lock and male luer lock to ensure safe and secure connection
  • Connecting blood tubing with dialysis machine
  • Ridged finger grip around the outside edge for better grip
  • Allowing only sterile air to pass through, protecting patients and equipment from cross contamination
  • Hydrophobic barrier separating blood and dialyzing fluid from sensitive transducer devices
Transducer-protectors 900 X 600

Specifications

Filter Media 0.22μm Hydrophobic PTFE
Housing Medical grade ABS
Dimensions 25.5*21.2mm (Luer connector included)
Weight 2.2g
Venting area 1.13cm²
Air flow rate 0.22μm ≥25ml/min@5kpa
Max burst pressure 4.0bar (58.0psi)
Operating temperature 5-40℃
Liquid burst pressure 0.22μm ≥3.0bar (43.5psi)
Connector Luer lock connector compatible with GB/T 1962.2-2001(ISO 594-2).
Pyrogenicity ≤0.5EU/PC
Storage temperature 0-50℃
Sterilization Ethylene oxide (Max 60℃)

Packing Details

6000pcs/carton containing 3 inner PE bags with 2000 pcs each. Each bag is sealed individually with protective air cushion between each other. Each bag is labeled with necessary information, and different lots in one carton are wrapped and labeled separately to prevent mix up.

Frequently Asked Questions

What is the primary function of the Transducer Protector?

It serves as an inline protective medical device for pressure monitors within the hemodialysis system, preventing contamination and protecting patients and equipment from cross-contamination.

What filter media is used in the device?

The device uses a 0.22μm Hydrophobic PTFE membrane filter which allows only sterile air to pass through while acting as a barrier to keep blood and dialyzing fluid separated.

What type of connector does the Transducer Protector use?

It features female luer lock and male luer lock connectors compatible with GB/T 1962.2-2001 (ISO 594-2) to ensure a safe and secure connection.

What are the dimensions and weight of the product?

The dimensions are 25.5*21.2mm (including the luer connector) and it weighs approximately 2.2g.

How is this product packaged for bulk shipment?

It is packaged with 6000pcs per carton, containing 3 inner PE bags with 2000pcs each. Each bag is sealed individually with protective air cushions and labeled separately by lot.

What is the sterilization method used for the Transducer Protector?

The Transducer Protector is sterilized using Ethylene oxide (with a maximum temperature of 60℃).